A Preliminary study of the measurement of slice-width dose profiles (SWDP) on diagnostic x-ray CT scanners using PAGAT polymer gel dosimeters with optical CT read-out

Sarvenaz Sarabipour, Stephen Bosi, Brendan Hill, Clive Baldock

Research output: Contribution to journalConference paperpeer-review

5 Citations (Scopus)

Abstract

The slice-width dose profile (SWDP) is a measurement undertaken during acceptance testing and subsequent quality assurance measurements of diagnostic x-ray CT scanners for the determination, of patient dose. In a previous study (Hill B, Venning A J and Baldock C 2005 Med. Phys. 32 1589-1597) normoxic polymer gel dosimeters were used to measure dose, the SWDP and subsequently calculate computer tomography dose index (CTDI) during acceptance testing of a CT scanner. In the current study, a preliminary investigation was undertaken to determine the SWDP of a diagnostic x-ray CT scanner using the PAGAT polymer gel dosimeters with optical CT read-out.
Original languageEnglish
Pages (from-to)280-282
Number of pages3
JournalJournal of Physics: Conference Series
Volume56
DOIs
Publication statusPublished - 2006
Externally publishedYes
EventInternational Conference on Radiotherapy Gel Dosimetry (4th : 2006) - Sherbrooke, Quebec, Canada
Duration: 7 Aug 200610 Aug 2006

Fingerprint

Dive into the research topics of 'A Preliminary study of the measurement of slice-width dose profiles (SWDP) on diagnostic x-ray CT scanners using PAGAT polymer gel dosimeters with optical CT read-out'. Together they form a unique fingerprint.

Cite this