Abstract
Results are presented showing the volume dependence of the measured activity of 125I in Capintec radionuclide calibrators. Errors of up to 76% were encountered for commonly used syringes and vials when the manufacturer's calibration factor was used. Calibration factors have been determined for syringes and vials which reduce errors to less than 2%.
Original language | English |
---|---|
Pages (from-to) | 445-450 |
Number of pages | 6 |
Journal | Nuclear Medicine Communications |
Volume | 12 |
Issue number | 5 |
DOIs | |
Publication status | Published - 1991 |
Externally published | Yes |