Application of monomer/polymer gel dosimetry to study the effects of tissue inhomogeneities on intensity-modulated radiation therapy (IMRT) dose distributions

Koen Vergote*, Yves De Deene, Filip Claus, Werner De Gersem, Bart Van Duyse, Leen Paelinck, Eric Achten, Wilfried De Neve, Carlos De Wagter

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

41 Citations (Scopus)

Abstract

Background and purpose: When planning an intensity-modulated radiation therapy (IMRT) treatment in a heterogeneous region (e.g. the thorax), the dose computation algorithm of a treatment planning system may need to account for these inhomogeneities in order to obtain a reliable prediction of the dose distribution. An accurate dose verification technique such as monomer/polymer gel dosimetry is suggested to verify the outcome of the planning system. Materials and methods: The effects of low-density structures: (a) on narrow high-energy (18 MV) photon beams; and (b) on a class-solution IMRT treatment delivered to a thorax phantom have been examined using gel dosimetry. The used phantom contained air cavities that could be filled with water to simulate a homogeneous or heterogeneous configuration. The IMRT treatment for centrally located lung tumors was delivered on both cases, and gel derived dose maps were compared with computations by both the GRATIS and Helax-TMS planning system. Results: Dose rebuildup due to electronic disequilibrium in a narrow photon beam is demonstrated. The gel measurements showed good agreement with diamond detector measurements. Agreement between measured IMRT dose maps and dose computations was demonstrated by several quantitative techniques. An underdosage of the planning target volume (PTV) was revealed. The homogeneity of the phantom had only a minor influence on the dose distribution in the PTV. An expansion of low-level isodoses in the lung volume was predicted by collapsed cone computations in the heterogeneous case. Conclusions: For the class-solution described, the dose in centrally located mediastinal tumors can be computed with sufficient accuracy, even when neglecting the lower lung density. Polymer gel dosimetry proved to be a valuable technique to verify dose calculation algorithms for IMRT in 3D in heterogeneous configurations.

Original languageEnglish
Pages (from-to)119-128
Number of pages10
JournalRadiotherapy and Oncology
Volume67
Issue number1
DOIs
Publication statusPublished - 1 Apr 2003
Externally publishedYes

Keywords

  • Class-solution
  • Inhomogeneity
  • Intensity-modulated radiation therapy (IMRT)
  • Lung cancer
  • Monomer/polymer gel dosimetry