Abstract
The Beer-Lambert law (Hollas 2004; Atkins and Friedman 2005; Atkins and de Paula 2006)
A = εCl
is arguably the single most useful spectroscopy equation. It follows from considering what happens to a beam of photons propagating along the X direction as it passes through a uniform sample of absorbing molecules. If light of intensity I(X) enters a slice of the sample at position X, then the intensity I(X + δ X) after traveling an infinitesimal distance δX is
where Π is the probability that a photon that hits a molecule is absorbed. Now
number of molecules hit by a photon
(mathematical equation)
where N slice is the number of molecules in the slice of the sample of width δ X, σ molecule is the cross-sectional area of the molecules facing the incident light beam, σ molar is the corresponding molar quantity, and c is the concentration of the analyte.
The Beer-Lambert law is not valid if too few photons are emitted from the sample for the photomultiplier tube to count them accurately. It also breaks down when the sample is inhomogeneous or if there are concentration-dependent intermolecular interactions that affect the spectroscopy.
A = εCl
is arguably the single most useful spectroscopy equation. It follows from considering what happens to a beam of photons propagating along the X direction as it passes through a uniform sample of absorbing molecules. If light of intensity I(X) enters a slice of the sample at position X, then the intensity I(X + δ X) after traveling an infinitesimal distance δX is
I(X + d X) = I(X) –
(number of molecules hit by a photon) × Π
number of molecules hit by a photon
(mathematical equation)
where N slice is the number of molecules in the slice of the sample of width δ X, σ molecule is the cross-sectional area of the molecules facing the incident light beam, σ molar is the corresponding molar quantity, and c is the concentration of the analyte.
The Beer-Lambert law is not valid if too few photons are emitted from the sample for the photomultiplier tube to count them accurately. It also breaks down when the sample is inhomogeneous or if there are concentration-dependent intermolecular interactions that affect the spectroscopy.
Fingerprint
Dive into the research topics of 'Beer-Lambert law derivation'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver