Cochlear implant soft failures consensus development conference statement

Thomas J. Balkany*, A. V. Hodges, C. A. Buchman, W. M. Luxford, C. H. Pillsbury, P. S. Roland, J. K. Shallop, D. D. Backous, D. Franz, J. M. Graham, B. Hirsch, M. Luntz, J. K. Niparko, J. Patrick, S. L. Payne, F. F. Telischi, E. A. Tobey, E. Truy, S. Staller

*Corresponding author for this work

    Research output: Contribution to journalReview articlepeer-review

    105 Citations (Scopus)

    Abstract

    In summary, CI soft failure is an uncommon occurrence in which a device malfunction is suspected but cannot be proven. It is a working diagnosis, based on characteristic symptoms such as shocking sensations, popping sounds, intermittency, or unexplained progressive decrement in performance. Complete otological evaluation, CT scan, expert re-programming, exchange of all external hardware, and integrity testing by the manufacturer are necessary before the working diagnosis is made. The usual management is device replacement when aversive symptoms are intolerable or performance becomes unacceptable as determined by the team and family.

    Original languageEnglish
    Pages (from-to)815-818
    Number of pages4
    JournalOtology and Neurotology
    Volume26
    Issue number4
    DOIs
    Publication statusPublished - Jul 2005

    Fingerprint

    Dive into the research topics of 'Cochlear implant soft failures consensus development conference statement'. Together they form a unique fingerprint.

    Cite this