Abstract
DPOAE (2 f1 - f2) phase was measured across a 3-octave frequency range from two groups of newborns using ER10B+ and ER10C probe microphones. A marked phase shift was noted in the mid-to-high frequency range for newborn data recorded with the ER10C only. In contrast, the ER10B+ produced phase that was approximately invariant as a function of frequency for most of the range. Probe-related phase shifts can be effectively eliminated by correcting for variations in the phases of the primary tones. Results highlight the importance of detecting and correcting for system-related phase shifts so they are not misinterpreted as cochlear in origin.
| Original language | English |
|---|---|
| Pages (from-to) | EL49-EL55 |
| Number of pages | 7 |
| Journal | Journal of the Acoustical Society of America |
| Volume | 128 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - Jul 2010 |
| Externally published | Yes |