Abstract
Purpose: To summarize the results of endovascular abdominal aortic aneurysm (AAA) treatment using several endograft designs over a 4.5-year experience and offer comparisons on the various devices. Methods: From May 1992 to August 1996, 121 AAA patients meeting the criteria for an endoluminal repair were treated with 1 of 5 endograft designs in three configurations. The endografts were implanted in the operating room under fluoroscopic control. Follow-up included contrast-enhanced computed tomography within 10 days of operation, 6 months postoperatively, and annually thereafter. Results: Endografts were successfully deployed in 106 patients (88%). Fifteen cases were converted to open repair. Six procedure-related deaths occurred within 30 days owing to myocardial infarction (3), combined renal failure and septicemia (2), and multisystem failure (1). There were 36 local/vascular complications (30%) and 18 systemic/remote complications (15%). Of the 121 patients undergoing endoluminal AAA repair, 93 (77%) are currently alive and well with their AAAs excluded from the circulation. Conclusions: Trends in endoluminal AAA repair and prosthetic design point toward simpler devices and earlier treatment of smaller aneurysms once the long-term outcome of aortic endografting has been determined.
Original language | English |
---|---|
Pages (from-to) | 147-151 |
Number of pages | 5 |
Journal | Journal of Endovascular Surgery |
Volume | 4 |
Issue number | 2 |
DOIs | |
Publication status | Published - 1997 |
Keywords
- Complications
- Endografts
- Endoleaks
- Endovascular grafting
- Outcome