Abstract
Chloroform in the FlexyDos3D dosimeter acts as a radical initiator which brings about the colour change of the dosimeter when the radicals react with the leucomalachite green. However, the volatility of the chloroform likely results in a rapid loss of chloroform from the dosimeter. Gravimetric analysis and NMR diffusion-ordered coherence spectroscopy were used to measure the diffusion and evaporation rates of chloroform from the dosimeter and found that both rates were both significantly large resulting in a rapid loss of chloroform. Dose maps of irradiated phantoms aged for different times found a significant difference in the dose measurements of the dosimeter, likely a result of the different chloroform concentrations within the dosimeters.
Original language | English |
---|---|
Article number | 012044 |
Pages (from-to) | 1-6 |
Number of pages | 6 |
Journal | Journal of Physics: Conference Series |
Volume | 1305 |
Issue number | 1 |
DOIs | |
Publication status | Published - 2019 |
Event | 10th International Conference on 3D Radiation Dosimetry - Kunshan, China Duration: 16 Sept 2018 → 19 Sept 2018 |