Abstract
The reference air kerma rates of a set of individual ¹²⁵I seeds were calculated from current measurements of a calibrated re-entrant ionisation chamber. Single seeds were distributed to seven Australian brachytherapy centres for the same measurement with the user's instrumentation. Results are expressed as the ratio of the reference air kerma rate measured by the Australian Nuclear Science & Technology Organisation (ANSTO) to the reference air kerma rate measured at the centre. The intercomparison ratios of all participants were within ±5% of unity.
Original language | English |
---|---|
Pages (from-to) | 517-518 |
Number of pages | 2 |
Journal | Applied Radiation and Isotopes |
Volume | 65 |
Issue number | 5 |
DOIs | |
Publication status | Published - 2007 |
Externally published | Yes |