Abstract
This study has investigated the relationship between linear attenuation coefficients (μ) and Hounsfield units (HUs) for six materials covering the range of values found clinically. Narrow-beam μ values were measured by performing radionuclide transmission scans using 99mTc, 123I, 131I, 201Tl and 111In. The μ values were compared to published data. The relationships between μ and HU were determined. These relationships can be used to convert computed tomography (CT) images to μ-maps for single photon emission computed tomography (SPECT) attenuation correction.
| Original language | English |
|---|---|
| Pages (from-to) | 1206-1212 |
| Number of pages | 7 |
| Journal | Applied Radiation and Isotopes |
| Volume | 66 |
| Issue number | 9 |
| DOIs | |
| Publication status | Published - 2008 |
| Externally published | Yes |
Keywords
- Attenuation correction
- CT
- Radionuclide
- SPECT