Abstract
Background: Patients with a cardiac implantable electronic device (CIED)-induced tricuspid regurgitation (TR) have an increased mortality and morbidity. However, the impact of CIED-lead extraction and its indications are not well-defined. Case summary: A 69-year-old woman presented with recurrent hospital admissions for right heart failure refractory to medical therapy, on the background of a single-chamber permanent pacemaker (Biotronik) implanted 6 years ago for tachycardia-bradycardia syndrome. Transoesophageal echocardiography identified severe TR which was predominantly CIED-induced from a lead impingement of the posterior tricuspid valve (TV) leaflet preventing adequate leaflet coaptation. This had progressed to cause a degree of secondary functional TR. The patient underwent pacing lead extraction followed by epicardial lead placement via minithoracotomy, with significant symptomatic and echographic improvement of TR. Discussion: CIED-induced TR from a lead impingement of TV leaflets carries the highest risk of TR and its consequences. This case illustrates the significance of the relationship between CIED-leads and the TV, which impacts management strategy. We recommend a mechanistic approach and incorporating CIED-lead interaction with the TV apparatus as the underlying principle in developing future management guidelines for CIED-induced TR.
Original language | English |
---|---|
Article number | ytac170 |
Pages (from-to) | 1-6 |
Number of pages | 6 |
Journal | European Heart Journal - Case Reports |
Volume | 6 |
Issue number | 7 |
DOIs | |
Publication status | Published - 6 Jul 2022 |
Bibliographical note
Copyright the Author(s) 2022. Version archived for private and non-commercial use with the permission of the author/s and according to publisher conditions. For further rights please contact the publisher.Keywords
- Cardiac implantable electronic devices
- Case report
- Lead impingement
- Pacing lead extraction
- Tricuspid regurgitation