PRESAGE® has previously shown potential for 3D dosimetry of heavy particles. A new formulation has specifically developed for dosimetry of protons/heavy ions. This work provides a preliminary characterization the new formulation of PRESAGE® by measuring optical absorbance and dose response after irradiating by a 62 MeV proton beam for a dose range of 0.5 - 20 Gy. Results show linear dose response and the evolution over time of the optical density of the 3D dosimeter.
| Original language | English |
|---|---|
| Pages (from-to) | 012058-1-012058-4 |
| Number of pages | 4 |
| Journal | Journal of Physics: Conference Series |
| Volume | 444 |
| DOIs | |
| Publication status | Published - 2013 |
| Event | International Conference on 3D Radiation Dosimetry (7th : 2012) - Sydney Duration: 4 Dec 2012 → 8 Dec 2012 |
Bibliographical note
Copyright the author(s). This paper was published in Journal of physics: conference series, Vol. 444, 012058. The original article can be found at http://dx.doi.org/10.1088/1742-6596/444/1/012058. Version archived for private and non-commercial use with the permission of the author/s and according to publisher conditions. For further rights please contact the publisher.Fingerprint