Abstract
Audible squeaking in total hip replacements with ceramic-on-ceramic bearings is a rare problem. Acetabular component orientation was compared for 17 squeaking hips and 17 matched controls. Ninety-four percent of control hips were in a range of 25° ± 10° anteversion and 45° ± 10° inclination, but only 35% of squeaking hips were in this range (P = .0003). Eight hips squeak with bending. Four hips squeak with walking, and 5 hips squeak after prolonged periods of walking. Hips that squeaked with walking had acetabular components that were more anteverted (40°) than hips that squeaked with bending (19°) (P = .001) or prolonged walking (18°) (P = .020). The hips started squeaking after an average of 14 months. Patients with squeaking hips were younger, heavier, and taller than patients with silent hips.
| Original language | English |
|---|---|
| Pages (from-to) | 496-503 |
| Number of pages | 8 |
| Journal | Journal of Arthroplasty |
| Volume | 22 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - Jun 2007 |
| Externally published | Yes |
Keywords
- alumina
- ceramic
- hip
- squeak
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver