TY - GEN
T1 - Stress sensing in pedicle screw
T2 - 18th International Conference on Sensing Technology, ICST 2025
AU - Kar, Ankita
AU - George, Boby
AU - Gooneratne, Chinthaka Pasan
AU - Kailash, K. karthik
AU - Mukhopadhyay, Subhas C.
PY - 2025
Y1 - 2025
N2 - Pedicle screw fixation is a widely adopted technique for spinal stabilization, yet mechanical failure due to excessive stress concentration remains a significant concern in long-term clinical outcomes. This study investigates the stress distribution within a functional pedicle screw under physiologically relevant loads and explores the feasibility of integrating real-time sensing mechanisms for in vivo monitoring. A simulation-based approach was employed to explore potential sensor placement locations, aiming to capture representative stress patterns while ensuring the structural integrity of the screw. Finite element analysis was performed under a range of physiological boundary conditions; however, a reference axial load of 500 N (51kg) was selected for a detailed evaluation. The findings highlight critical stress concentration zones and propose optimal sensor placement regions, thereby offering insights into the design of smart pedicle screws capable of continuous load monitoring. This work contributes to the advancement of smart spinal implants that can improve postoperative assessment and improve patient outcomes through early detection of implant overload.
AB - Pedicle screw fixation is a widely adopted technique for spinal stabilization, yet mechanical failure due to excessive stress concentration remains a significant concern in long-term clinical outcomes. This study investigates the stress distribution within a functional pedicle screw under physiologically relevant loads and explores the feasibility of integrating real-time sensing mechanisms for in vivo monitoring. A simulation-based approach was employed to explore potential sensor placement locations, aiming to capture representative stress patterns while ensuring the structural integrity of the screw. Finite element analysis was performed under a range of physiological boundary conditions; however, a reference axial load of 500 N (51kg) was selected for a detailed evaluation. The findings highlight critical stress concentration zones and propose optimal sensor placement regions, thereby offering insights into the design of smart pedicle screws capable of continuous load monitoring. This work contributes to the advancement of smart spinal implants that can improve postoperative assessment and improve patient outcomes through early detection of implant overload.
UR - https://www.scopus.com/pages/publications/105041278611
U2 - 10.1109/ICST66402.2025.11512408
DO - 10.1109/ICST66402.2025.11512408
M3 - Conference proceeding contribution
AN - SCOPUS:105041278611
SN - 9798331553722
SP - 298
EP - 302
BT - 2025 18th International Conference on Sensing Technology (ICST)
PB - Institute of Electrical and Electronics Engineers (IEEE)
CY - Piscataway, NJ
Y2 - 1 December 2025 through 3 December 2025
ER -