Temperature dependence on the dose response of the Fricke-gelatin-xylenol orange gel dosimeter

J. B. Davies, C. Baldock

Research output: Contribution to journalArticlepeer-review

18 Citations (Scopus)

Abstract

The effect of temperature during optical absorbance measurements and irradiation on the values of the molar absorption coefficient, (εm)t, and radiation chemical yield, Gt′, respectively, for the Fricke–gelatin–xylenol orange (FGX) gel dosimeter were investigated. At 20 °C, the product (εm)tGt′ was found to be 6.76×10⁻³ m² J⁻¹. While an irradiation temperature coefficient, k₂, was evident (−0.53% °C⁻¹), no temperature effect was found during measurement in the range 16–25 °C.
Original languageEnglish
Pages (from-to)660-662
Number of pages3
JournalRadiation Physics and Chemistry
Volume79
Issue number5
DOIs
Publication statusPublished - 2010
Externally publishedYes

Keywords

  • Dosimetry
  • Extinction coefficient
  • Fricke gel
  • G-value
  • Temperature coefficient